beautypg.com

1 pulse input channels – Campbell Scientific CR200/CR200X-series Dataloggers User Manual

Page 55

background image

Section 4. Sensor Support

Note: The PulseCount () instruction should not be used in a conditional

statement or subroutine. To ensure all pulses are detected, it must be executed

each scan.

Execution of PulseCount () within a scan involves determining the accumulated

counts in each dedicated 16-bit counter since execution of the last PulseCount

(). PulseCount () parameter (POption) determines if the output is in counts

(POption = 0) or frequency (POption = 1). Counts are the preferred output

option for measuring number of tips from a tipping bucket rain gage, or the

number of times a door opens. Many pulse sensors, such as anemometers and

flow meters, are calibrated in terms of frequency (Hz or counts / second), and

are usually measured with the frequency option.

Resolution of the pulse counters is one count. Resolution of frequency is

(1/scan interval). For example, the frequency resolution of PulseCount()

returning a result every 1 second is 1 Hz. The resultant measurement will

bounce around by the resolution. For example, if you are scanning a 2.5 Hz

input once a second, in some intervals there will be 2 counts and in some 3. If

the pulse measurement is averaged, the correct value will be the result.

Accuracy is limited by a small scan interval error of ±(3 ppm of scan interval +

10 µs) plus the measurement resolution error of ± 1 Hz. The sum is essentially ±

1 Hz. Extending a 1 second measurement interval to 10 seconds, either by

increasing the scan interval or by averaging, improves the resulting frequency

resolution from 1 Hz to 0.1 Hz. Averaging can be accomplished by the Average

() instruction or by computing a running or spatial average through

programming.

4.4.1 Pulse input Channels

Read More! Review

pulse counter specifications

p. 35. Review pulse counter

programming in CRBASIC Help for the PulseCount () instruction.

43