beautypg.com

3 switched unregulated (nominal 12 volt), 2 voltage measurement, 1 measurement sequence – Campbell Scientific CR200/CR200X-series Dataloggers User Manual

Page 50: 2 measurement accuracy, Table 4. current sourcing limits

background image

Section 4. Sensor Support

4.1.3 Switched Unregulated (Nominal 12 Volt)

Voltage on the SW Battery terminal will change with CR200(X) supply voltage.

The CRBASIC instruction SWBatt () controls the SW Battery terminal.

Table 4. Current Sourcing Limits

Terminal

Limit

VX1, VX2

25 mA @ 2.5V
10 mA @ 5V

SW Battery

< 900 mA @ 20°C

< 729 mA @ 40°C

< 630 mA @ 50°C

< 567 mA @ 60°C

< 400 mA @ 85°C

4.2 Voltage Measurement

4.2.1 Measurement Sequence

The first step in a voltage measurement is a calibration to measure the ground

offset. The calibration is performed once for each voltage measurement. The

CR200(X) measures analog voltages with a sample and hold analog to digital

(A/D) conversion. The A/D conversion is made with a 12-bit successive

approximation technique which resolves the signal voltage to one part in 4096

of the full scale 2.5 V range, which is 0.6 millivolts.

To reduce noise, 10 measurements are rapidly made and averaged to form the

result returned. The measurements that go into the average each take about 26

microseconds.

4.2.2 Measurement Accuracy

CR200(X) analog measurement error is calculated as

Error = Gain Error (%) + Offset Error

Gain error is expressed as ±% and is a function of input voltage and CR200(X)

temperature. It increases with component temperature and aging. Between -

40°C and +50°C, gain error is typically ±0.25% of the reading with a 1.2

millivolt offset. Worst case over that same temperature range is ±1% of the

reading with a 2.4 millivolt offset.

FIGURE. Voltage Measurement Accuracy (-40° to +50°C)

p. 40 illustrates that

as magnitude of input voltage decreases, measurement error decreases.

38