Section 4. Sensor Support

Several features give the CR200(X) the flexibility to measure many sensor

types. Contact a Campbell Scientific applications engineer if assistance is

required to assess sensor compatibility.

4.1 Powering Sensors

Read More! See

CR200(X) Power Supply

(p. 53).

The CR200(X) is a convenient router of power for sensors and peripherals

requiring a 2.5, 5 or 12 VDC source. It has one program-controlled switched 12

Volt terminal (SW Battery), and two switched voltage excitation terminals

(VX1, VX2). These terminals limit current internally for protection against

accidental short circuits. Sensors can be powered continuously by connecting to

the Battery + terminal, which is the same as connecting sensors directly to the

external battery.

Voltage on the Battery + and SW Battery terminals will changed with the DC

supply used to power the CR200(X). The switched voltage excitation terminals

are typically used to supply a brief excitation for a bridge measurement with the

ExDelSE () instruction, but can also be programmed to provide a continuous 2.5

V or 5 V using the Excite () instruction, which is enough to power many

sensors.

Switched Unregulated (Nominal 12 Volt)

(p. 38) shows the current

limits of SW Battery. Greatly reduced output voltages associated with SW

Battery, VX1, and VX2 due to current limit may occur if the current limits

given in the table in

TABLE. Current Sourcing Limits

p. 38 are exceeded.

4.1.1 Switched Precision

Voltage (0, 2500, 5000 mV)

Two switched analog output (excitation) terminals, VX1 and VX2, operate

under program control. Check the accuracy specification of these channels in

Specifications

(p. 35) to understand their limitations. Specifications are only

applicable for loads not exceeding ±25 mA for 2.5 V excitation, or 10 mA for 5

V excitation. CRBASIC instructions that control excitation channels include:

ExDelSE ()

ExciteV ()

4.1.2 Continuous Unregulated (Nominal 12 Volt)

Voltage on the Battery + terminal will change with the CR200(X) external

battery voltage.

37