4 replacing the internal battery, Figure 52: enclosure – Campbell Scientific CR200/CR200X-series Dataloggers User Manual
Page 160

Section 17. Care and Maintenance
Figure 52: Enclosure
17.4 Replacing the Internal Battery
Caution Fire, explosion, and severe burn hazard! Misuse or improper
installation of the lithium battery can cause severe injury. Do not
recharge, disassemble, heat above 100°C (212°F), solder directly to the
cell, incinerate, or expose contents to water. Dispose of spent lithium
batteries properly.
The CR200(X) contains a lithium battery that operates the clock and SRAM
when the CR200(X) is not powered. The CR200(X) does not draw power from
the lithium battery while it is powered by an external source. In a CR200(X)
stored at room temperature, the lithium battery should last approximately 5
years (less at temperature extremes). Where the CR200(X) is powered most or
all of the time the lithium cell should last much longer.
The internal lithium battery must be replaced periodically. Factory replacement
is recommended. Contact Campbell Scientific to obtain an RMA prior to
shipping the CR200(X).
If the lithium cell is removed or allowed to discharge below the safe level, the
CR200(X) will still operate correctly while powered. Without the lithium
battery, the clock will reset and data is lost when power is removed.
148