Section 17. Care and Maintenance

Figure 52: Enclosure

17.4 Replacing the Internal Battery

Caution Fire, explosion, and severe burn hazard! Misuse or improper

installation of the lithium battery can cause severe injury. Do not

recharge, disassemble, heat above 100°C (212°F), solder directly to the

cell, incinerate, or expose contents to water. Dispose of spent lithium

batteries properly.

The CR200(X) contains a lithium battery that operates the clock and SRAM

when the CR200(X) is not powered. The CR200(X) does not draw power from

the lithium battery while it is powered by an external source. In a CR200(X)

stored at room temperature, the lithium battery should last approximately 5

years (less at temperature extremes). Where the CR200(X) is powered most or

all of the time the lithium cell should last much longer.

The internal lithium battery must be replaced periodically. Factory replacement

is recommended. Contact Campbell Scientific to obtain an RMA prior to

shipping the CR200(X).

If the lithium cell is removed or allowed to discharge below the safe level, the

CR200(X) will still operate correctly while powered. Without the lithium

battery, the clock will reset and data is lost when power is removed.