beautypg.com

Section 12. memory and data storage, 1 data storage, 1 data table storage – Campbell Scientific CR200/CR200X-series Dataloggers User Manual

Page 141

background image

Section 12. Memory and Data Storage

12.1 Data Storage

The CR200(X) can be programmed to store each measurement or, more

commonly, to store processed values such as averages, maxima, minima, etc.

Data are stored periodically or conditionally in data tables as directed by the

CRBASIC program (

CRBASIC EXAMPLE. Proper Program Structure

p. 73).

The DataTable () instruction allows the user to set the size of the data table. The

maximum number of tables that can be created by the program is 4 for CR200

series dataloggers and 8 for CR200(X) series dataloggers.

12.1.1 Data Table Storage

Data tables are stored in Serial Flash EEPROM. Data remain in memory when

the CR200(X) is powered down. Data are erased when a different program is

loaded and run. The Serial Flash EEPROM is good for more than 100,000 write

cycles.

Caution! If an EEPROM memory failure is detected, the datalogger

suspends running the program and red LED flashes twice every scan

interval. “Trap Code” in the Status Table will be set to 16. The

datalogger must be returned to CSI to replace the Serial Flash

EEPROM.

The CR200(X) stores the final storage data in the Serial Flash EEPROM. The

Serial Flash EEPROM also is the storage area for a file called the Table

Definition File (TDF). When a CRBASIC program is downloaded to the

CR200(X), the TDF is extracted from the compiled version of the CRBASIC

program and stored in the Serial Flash EEPROM. Any remaining memory can

be used for the final storage allocation. In the CR200, there are 128k bytes or

512k bytes of Serial Flash EEPROM (dataloggers with a 512 kbyte EEPROM

will have 512K on their label). Up to 5.12k bytes of that memory can be used

for the TDF (if the TDF is greater than the maximum, the compile will fail). If

the data table allocation or size of all or any of the tables use too much memory,

there will NOT be a compile error. The CR200(X) will force the allocation to

work by adding all the sizes for each table, checking if the sum is greater than

the Serial Flash EEPROM and dropping the allocation for each table by 90

percent. This is repeated until the sums of the allocations are below the Serial

Flash EEPROM size.

129