RATE SCALING EXAMPLES (Cont’d)

We now have all the information necessary to scale the rate. The Total Scaling

Factor “K

T

”, is calculated using Formula #1 as shown below.

K

T

= Display Units/Number of Pulses

K

T

= 810/5600 = 0.144643

It is noticed that there are more significant digits in the Total Scale Factor,

“K

T

”, than there are available for Scale Factor, “SF”. In order to acquire the

maximum amount of significant digits for the Scale Factor, Formula #2 is used

and a Scale Multiplier value of 0.1 is selected.

SF = K

T

/SCM

SF = 0.144643/0.1 = 1.4464

The Scale Factor, “SF”, then becomes 1.4464 (1.44643 rounded to 4 decimal

places). This provides the maximum amount of conversion accuracy possible.

The final step is to select a Rate Conversion Factor. The Rate Conversion

Factor is chosen to be Rate Per Minute as was required.