PROGRAMMING THE GEMINI 5200 IN DUAL RATE WITHOUT DISPLAY C MODE

The rate indicators use a time interval method (1/tau) to calculate the rate

value. This method enables high resolution at all input rates. The unit counts

input pulses and after a programmable minimum update time has occurred, it

waits until the next count edge occurs, takes the elapsed time and number of

edges, and calculates the rate value. At slower rates, averaging can be

accomplished by programming the “Minimum Update Time” (0.5 sec. to 16 sec.)

for the desired response. Extensive scaling capabilities allow practically any

desired reading at very slow count rates.

In the Dual Rate, without Display C mode, the two rate indicators operate in an

independent non-synchronized fashion. Input A serves as the input to the Rate A

indicator and Input B serves as the input to the Rate B indicator. All modes are

independently selected for each rate channel.

CODE 41 - UNIT PERSONALITY

The unit personality mode is the first function code that is programmed. This

mode will determine the basic operational characteristics of the unit. This

programming section deals with the unit programmed as a Dual Rate Indicator

without Display C, mode 1. The “Programming The A/B, (A-B)/B, And A-B

Display C Modes” section covers programming of the Gemini 5200 when in Unit

Personality modes 2, 3, and 4.

Once the Unit Personality mode is changed, the factory settings will

automatically be loaded into the unit’s memory. This provides a known starting

place. If at any point, it is desired to return to the factory settings for the current

mode, the unit personality is called up, a minus is put in front of the mode by

pressing the “+/-” key, and the “E” key is pushed to load the factory settings. The

factory settings are listed on page 5 and 6.
[41 1] DUAL RATE INDICATION - In this mode, the Gemini 5200 operates as

two independent rate indicators.

[41 2] DUAL RATE WITH A/B DISPLAY C - In this unit personality, the

Gemini 5200 operates as a dual rate indicator with an A/B display C function.

See “Programming The A/B, (A-B)/B, And A-B Display C Modes” section for

programming details.

[41 3] DUAL RATE WITH A-B DISPLAY C - In this unit personality, the

Gemini 5200 operates as a dual rate indicator with an (A-B) Display C

function. See “Programming The A/B, (A-B)/B, And A-B Display C Modes”

section for programming details.

[41 4] DUAL RATE WITH (A-B)/B DISPLAY C - In this unit personality, the

Gemini 5200 operates as a dual rate indicator with an (A-B)/B Display C

function. See “Programming The A/B, (A-B)/B, And A-B Display C Modes”

section for programming details.

[41 -1]

[41 -2]

A minus indicator preceding the mode will cause

[41 -3]

the factory settings to be loaded when entered.

[41 -4]

CODE 42 - RESET KEY & TERMINAL ACTUATION MODES

The Reset Button & Terminal Actuation Modes controls the effect that the

reset button and terminal have on the two display channels. Resetting will not

affect the rate displays or update period in any manner. Reset only affects the

output(s) associated with the rate channels.

There is a separate “Rst A” terminal provided which makes it possible to

independently reset both the Rate A and B Outputs.
[42 1] RESET RATE A OUTPUT(S)

[42 2] RESET RATE B OUTPUT(S)

[42 3] RESET RATE A & B OUTPUTS

CODE 43 - RATE B CONVERSION FACTOR

The Rate B Conversion Factor is used to convert Display B to indicate rate in

the desired time unit. See Appendix “F” - Scaling For Rate And Display C.
[43 1] Rate Per Second (x1)

[43 2] Rate Per Minute (x60)

[43 3] Rate Per Hour (x3600)

-9-