10. TROUBLESHOOTING

If the communication is unavailable, check the following items.

†

Whether all devices related to communication are turned on.

†

Whether connections are correct.

†

Whether the number of connected instruments and connection distance are as specified.

†

Whether communication conditions coincide between the master station (host computer) and slave stations.

† Transmission

speed:

† 9600bps

† 19200bps
† 38400bps

† Data length:

8 bits

† Stop

bit:

1

bit

† Parity:

† odd

† even

† none

†

Whether send/receive signal timing conforms to Section 5.3 in this manual.

† Whether the station No. designated as send destination by the master station coincides with the station No. of

the connected transmitter.

†

Whether more than one instrument connected on the same transmission line shares the same station No.

†

Whether the station No. of instruments is set at other than 0.

If it is 0, the communication function does not work.

†

Whether the 5th digit of type code of this Flow transmitter is A or B ?