2 received signal (optional function) – Dwyer UXF3 User Manual
Page 63
- 59 -
9.14.2
Received Signal (optional function)
Click “RECEIVED SIGNAL”, and the following screen appears.
Fig. 9-22 < Received signal screen >
Select one from forward direction received wave, reverse direction received wave, forward direction filter, reverse
direction filter and correlation waveform. Depending on measurement method (method 1, method 2 and method
3), items which can be selected vary as shown below. Trigger level is also displayed.
Left-click the mouse while pressing the shift key to specify the screen range, and the selected range is magnified.
Press the R key to return to original status.
Ɣ Method 1: One from forward direction, reverse direction and correlation waveform can be selected.
Ɣ Method 2: One from forward direction, reverse direction, forward direction filter and reverse direction filter
can be selected.
Ɣ Method 3: One from forward direction, reverse direction, forward direction filter and reverse direction filter
can be selected.
[Start]...........................Starts reading.
[Stop] ...........................Stops reading.
[Save As CSV].............Saves the measurement result in a file in CSV format. Click the button, and you are
prompted to enter the name of a file to which the data is to be saved. Specify the
destination to save and enter the file name, and a CSV file is created.
Point
1. Startup is completed within 3 to 6 waves.
2. No peak (amplitude) fluctuation should be observed.
Otherwise air is mixed in.
Signal
wavefor
m
Trigger
level