Dwyer UXF3 User Manual
Page 49
- 45 -
[Setting] .......................Sends the setting of the selected item (check box set to ON (
;)), reflecting the response
value on the setting.
[READ] .......................Reads the setting of the selected item (check box set to ON (
;)), reflecting the response
value on the setting.
[Save] ..........................Reflects the setting sent by pressing the [Setting] button on the flow transmitter. Be sure to
press the [Save] button if the setting is changed.
[Check ON/OFF] .........Set the check box to ON to select all the items (to set all the check boxes to ON (
;)). Set
the check box to OFF (
) to release the selection of all the items (to set all the check boxes
to OFF (
)).
[CLEAR] .....................Clears zero calibration.
[SET ZERO]................Executes zero calibration adjustment.
Table 9-4 < Piping Specifications >
Item Content
OUTER DIAMETER
Enter in the range from 6.00 to 6200.00 mm (two decimal places) for metric
system, and from 0.2362 to 244.1000 inch (four decimal places) for inch system.
PIPE MATERIAL
Select from carbon steel, stainless steel, PVC, copper, cast iron, aluminum, FRP,
ductile iron, PEEK, PVDF, acrylic, PP, and pipe S.V.
PIPE SOUND VELOCITY
Enter in the range from 1000 to 3700 m/s (no decimal point) for metric system and
from 3280 to 12140 ft/s (no decimal point) for inch system. (If “Pipe S.V.” is
selected as piping material.)
WALL THICKNESS
Enter in the range from 0.10 to 100.00 mm (two decimal places) for metric system,
and from 0.0039 to 3.9380 inch (four decimal places) for inch system.
LINING MATERIAL
Select from no lining, tar epoxy, mortar, rubber, Teflon, pyrex glass, PVC and
lining S.V.
LINING SOUND VELOCITY
Enter in the range from 1000 to 3700 m/s (no decimal point) for metric system, and
from 3280 to 12140 ft/s (no decimal point) for inch system. (If “Lining S.V.” is
selected as lining material”.)
LINING THICKNESS
Enter in the range from 0.01 to 100.00 mm (two decimal places) for metric system,
and 0.0003 to 3.9380 inch (four decimal places) for inch system. (If “No lining”
is selected as lining material.)
KIND OF FLUID
Select for water, seawater, dist. water, ammonia, alcohol, benzene, bromide,
ethanol, glycol, kerosene, milk, methanol, toluene, lube oil, fuel oil, petrol, coolant
R410, and fluid S.V.
FLUID S.V.
Enter in the range from 300 to 2500 m/s (no decimal point) for metric system, and
from 984 to 8203 ft/s (no decimal point) for inch system. (If “Fluid S.V.” is
selected as fluid type”).
VISCOSITY
Enter in the range from 0.001 to 999.999 E·6 m
2
/s for metric system, and from
0.0107 to 10764 E·6 ft
2
/s for inch system.
SENSOR SPACING
[Read] only is valid.
Table 9-5 < System >
Item Content
SENSOR MOUNT
Select from Z method and V method.
TRANSMIT VOLTAGE
Select from 20, 40, 80 and 160Vpp.
SENSOR TYPE
Select from Reserved, SX1-A, SX1-B,
SX1-C, PSX2-A, SX2-A, and SX2-B.
Bulletin F-107-UXF3-S