Bulletin f-107-uxf3-s – Dwyer UXF3 User Manual
Page 31
- 27 -
Relative
address
Register
No.
Data
type
Parameter
Read-out data/Write-in data setting range
Remarks
00C6
40199 int
Decimal point position of 2nd row
0: *.******, 1: **.*****, 2: ***.****,
3: ****.***, 4: *****.**, 5: ******.*,
6: ********.
Write-in is not
permitted in case
the row is
“Velocity”, “+
Total pulse” or “
Total pulse”.
00C8 40201
00CA 40203
00CC 40205
00CE 40207
00D0 40209
int
Sensor
type
2:RSVD, 3:RSVD, 4:RSVD,
5:SX1-A, 6:SX1-B,
7:SX1-C, 8:RSVD,
9:PSX2-A, 10:SX2-A, 11:SX2-B
00D2 40211
long Outside
diameter
Metric system: 2 places after decimal point,
6.00 to 6200.00mm
English system: 4 places after decimal point,
0.2362 to 244.100inch
00D6 40215
int
Pipe
material
0: Carbon steel, 1: Stainless steel, 2: PVC,
3: Copper, 4: Cast iron, 5: Aluminum, 6: FRP,
7: Ductile iron, 8: PEEK, 9: PVDF, 10: Acrylic,
11: PP, 12: Pipe S.V.
00D8 40217
int
Pipe
S.V.
Metric system: Decimal point fixed, 1000 to
3700m/s
English system: Decimal point fixed, 3280 to
12140ft/s
Write-in is
permitted in
case pipe
material is “12:
Pipe S.V.”
00DA
40219 long
Pipe wall thickness
Metric system: 2 places after decimal point,
0.10 to 100.00mm
English system: 4 places after decimal point,
0.0039 to 3.9380inch
00DE 40223
int
Lining
material
0: No lining, 1: Tar epoxy, 2: Mortar, 3: Rubber,
4: Teflon, 5: Pyrex glass, 6: PVC, 7: Lining S.V.
00E0 40225
int
Lining
S.V.
Metric system: Decimal point fixed, 1000 to
3700m/s
English system: Decimal point fixed, 3280 to
12140ft/s
Write-in is
permitted in
case lining
material is “7:
Lining S.V.”
00E2 40227
long Lining
thickness
Metric system: 2 places after decimal point,
0.10 to 100.00mm
English system: 4 places after decimal point,
0.0003 to 3.9380inch
00E6
40231 int
Kind of fluid
0: Water, 1: Seawater, 2: Distilled water,
3: Ammonia, 4: Alcohol, 5: Benzene,
6: Bromide, 7: Ethanol, 8: Glycol, 9: Kerosene,
10: Milk, 11: Methanol, 12: Toluol, 13: Lube oil,
14: Fuel oil, 15: Petrol, 16: Refrigerant R410,
17: Fluid S.V.
00E8 40233
int
Fluid
S.V.
Metric system: Decimal point fixed, 300 to
2500m/s
English system: Decimal point fixed, 984 to
8203ft/s
00EA 40235
double Viscosity
Metric system: 32-bit floating data; 0.001 to
999.999 E-6m
2
/s
English system: 32-bit floating data; 0.0107 to
10764 E-6ft
2
/s
00F2
40243 int
Sensor mount
0: V method, 1: Z method
00F4
40245 int
Transmission voltage
0:20 Vpp, 1:40 Vpp, 2:80 Vpp, 3:160 Vpp
00F6 40247
int
Not
use
00F8 40249
int
Not
use
00FA 40251
int
Not
use
00FC 40253
int
Not
use
00FE 40255
int
Not
use
0100 40257
int
System
unit
*1
0: Metric, 1: English
0102 40259
int
System
language
0: English, 1: Japanese, 2: German, 3: French,
4: Spanish
0104
40261 int
ID No. setup
Decimal point fixed, 0 to 9999
Bulletin F-107-UXF3-S