Table 9-6 < Range Setting >

Item Content

RANGE UNIT

Select from L/s, L/min, L/h, L/d, kL/d, ML/d, m

3

/s, m

3

/min, m

3

/h, m

3

/d, k m

3

3/d, M m

3

/d,

BBL/s, BBL/min, BBL/h, BBL/d, kBBL/d, MBBL/d [ft

3

/s, ft

3

/min, ft

3

/h, ft

3

/d, kft/d,

Mft

3

/d, gal/s, gal/min, gal/h, gal/d, kgal/d, Mgal/d, BBL/s, BBL/min, BBL/h, BBL/d,

kBBL/d, MBBL/d]* Of which [ ]: unit is in case of inch system.

RANGE TYPE

Select from SINGLE, AUTO 2, BI-DIR, BI-DIR AUTO 2.

FULL SCALE 1

Enter 0, ±0.3 to 32 m/s fitting value (comply with range unit).

FULL SCALE 2

Enter 0, ±0.3 to 32 m/s fitting value (comply with range unit).

HYSTERISIS

Enter in the range of 0.00 to 20.00%. (2 places after decimal point)

OUTPUT LIMIT LOW

Enter in the range of -20 to 0%.

OUTPUT LIMIT HIGH

Enter in the range of 100 to 120%.

OUTPUT BURNOUT

Select from NOT USED, HOLD, UPPER, LOWER, ZERO.

BURNOUT TIMER

Enter in the range of 0 to 900sec.

RATE LIMIT

Enter 0 to 5 m/s fitting value (comply with range unit).

RATE LIMIT TIMER

Enter in the range of 0 to 900 sec.

Table 9-7 < Damping >

Item Content

DAMPING

Enter in the range of 0.0 to 100.0 sec. (1 place after decimal point)

Table 9-8 < Low Flow Rate Cut >

Item Content

CUT OFF

Enter 0 to 5 m/s fitting value (comply with range unit).

Table 9-9 < Output Correction >

Item Content

ZERO

Enter –5 to 5 m/s fitting value (comply with range unit).

SPAN

Enter in the range of ±200.00%. (2 places after decimal point)

Table 9-10 < Operation Mode >

Item Content

OPERATION MODE

Select from NORMAL, HIGH SPEED.