Dwyer UXF3 User Manual
Page 55
- 51 -
[Check ON/OFF] .........Set the check box to ON to select all the items (to set all the check boxes to ON (
;)). Set
the check box to OFF (
) to release the selection of all the items (to set all the check boxes
to OFF (
)).
Table 9-12 < Status output setting >
Item Content
Output
Select from Not use, + Total pulse, – Total pulse, Full scale 2, Alarm, Flow switch,
Total switch, AO range over, Pulse range over, and – Flow direction.
Alarm
Select from All, Hardware error, and Process error (when alarm is selected for DO1
output).
Flow rate switch
Select from Upper flow rate limit (Flow switch High) and Lower flow rate limit
(Flow switch Low) (when flow rate switch is selected for DO1 output).
Flow switch High
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
Flow switch Low
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
DO1
Total switch
Enter in the range from 0 to 99999999. (Use the same unit as the total unit.)
Output
Select from Not use, + Total pulse, – Total pulse, Full scale 2, Alarm, Flow switch,
Total switch, AO range over, Pulse range over, and – Flow direction.
Alarm
Select from All, Hardware error, and Process error (when alarm is selected for DO1
output).
Flow rate switch
Select from Upper flow rate limit (Flow switch High) and Lower flow rate limit
(Flow switch Low) (when flow rate switch is selected for DO1 output).
Upper flow rate limit
(Flow switch High)
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
Lower flow rate limit
(Flow switch Low)
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
DO2
Total switch
Enter in the range from 0 to 99999999. (Use the same unit as the total unit.)
Output
Select from Not use, + Total pulse, – Total pulse, Full scale 2, Alarm, Flow switch,
Total switch, AO range over, Pulse range over, and – Flow direction.
Alarm
Select from All, Hardware error, and Process error (when alarm is selected for DO1
output).
Flow rate switch
Select from Upper flow rate limit (Flow switch High) and Lower flow rate limit
(Flow switch Low) (when flow rate switch is selected for DO1 output).
Upper flow rate limit
(Flow switch High)
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
Lower flow rate limit
(Flow switch Low)
Enter in the range from 0 to 32 m/s or equivalent. (Use the same unit as the range
unit.)
DO3
Total switch
Enter in the range from 0 to 99999999. (Use the same unit as the total unit.)
DO1 contact operation
Select ON at operation or OFF at operation.
DO2 contact operation
Ditto
DO3 contact operation
Ditto
Table 9-13 < Status input setting >
Item Content
DO1 input
Select from Not use, Total reset, and Zero calibration.
DO2 contact operation
Select ON at operation or OFF at operation.
Bulletin F-107-UXF3-S