7.2 Address

Map

See the instruction manual attached to the main unit for details of functions and setting ranges of individual

parameters.

Data type unsigned char : Byte data without sign. This data is handled in byte units. One data per address

int

: Word data with sign. This data is handled in word units. One data per two addresses

unsigned int : Word data without sign. This data is handled in word units. One data per two

addresses

Long

: 2-word data with sign. This data is handled in 2-word units. One data per four

addresses

float

: Floating data. This data is handled in 2-word units. One data per four addresses

double

: Floating data. This data is handled in 4-word units. One data per eight addresses

7.2.1

Word data [Read-out/Write-in]: Function code [ 03H, 10H ]

Relative

address

Register

No.

Data

type

Parameter

Read-out data/Write-in data setting range

Remarks

4XXXX

0000

40001 int

Damping

1 place after the decimal point, 0.0 to 100.0sec

0002 40003

int

Not

use

0004

40005 int

Flow rate unit

Metric system : 0:L/s, 1:L/min, 2:L/h, 3:L/d,

4:kL/d, 5:ML/d, 6:m

3

/s, 7:m

3

/min, 8:m

3

/h,

9:m

3

/d, 10:km

3

/d, 11:Mm

3

/d, 12:BBL/s,

13:BBL/min, 14:BBL/h, 15:BBL/d,
16:kBBL/d, 17:MBBL/d

English system : 0:gal/s, 1:gal/min, 2:gal/h,

3:gal/d, 4:kgal/d, 5:Mgal/d, 6:ft

3

/s, 7:ft

3

/min,

8:ft

3

/h, 9:ft

3

/d, 10:kft

3

/d, 11:Mft

3

/d, 12:BBL/s,

13:BBL/min, 14:BBL/h, 15:BBL/d,
16:kBBL/d, 17:MBBL/d

0006 40007

int

Range

type

0: Single range, 1: Auto 2 range, 2: Forward-
reverse range, 3: Forward-reverse auto 2 range

0008

40009 double

Full scale 1

Metric system: 64-bit floating data; 0, ±0.3 to

±32m/s as converted to flow rate

English system: 64-bit floating data

Unit: Flow rate

0010

40017 double

Full scale 2

Metric system: 64-bit floating data; 0, ±0.3 to

±32m/s as converted to flow rate

English system: 64-bit floating data

Unit: Flow rate

0018

40025 int

Range hysteresis

2 places after the decimal point, 0.00 to 20.00%

001A

40027 int

Burnout

0: Not use, 1: Hold, 2: Upper, 3: Lower, 4: Zero

001C

40029 int

Burnout timer

Decimal point fixed, 0 to 900sec

001E

40031 int

Output limit low

Decimal point fixed,

20 to 0%

0020

40033 int

Output limit high

Decimal point fixed, 100 to 120%

0022

40035 int

Rate limit timer

Decimal point fixed, 0 to 900sec

0024 40037

double Rate

limit

Metric system: 64-bit floating data; 0 to 5m/s

as converted to flow rate

English system: 64-bit floating data

Unit: Flow rate

002C

40045 double

Low flow rate cut

Metric system: 64-bit floating data; 0 to 5m/s

as converted to flow rate

English system: 64-bit floating data

Unit: Flow rate

0034 40053

double Calibration

zero

Metric system: 64-bit floating data; ±5m/s as

converted to flow rate

English system: 64-bit floating data

Unit: Flow rate

003C

40061 int

Calibration span

2 places after the decimal point, 0.00 to 200.00%

003E

40063 int

Operation mode

0: Normal, 1: High speed response mode

0040 40065

int

Total

unit

*1

Metric system : 0:mL, 1:L, 2:m

3

, 3:km

3

,

4:Mm

3

, 5:mBBL, 6:BBL, 7:kBBL

English system : 0:gal, 1:kgal, 2:ft

3

, 3:kft

3

,

4:Mft

3

, 5:mBBL, 6:BBL, 7:kBBL, 8:ACRf

Bulletin F-107-UXF3-S