7.2 Address
Map
See the instruction manual attached to the main unit for details of functions and setting ranges of individual
parameters.
Data type unsigned char : Byte data without sign. This data is handled in byte units. One data per address
int
: Word data with sign. This data is handled in word units. One data per two addresses
unsigned int : Word data without sign. This data is handled in word units. One data per two
addresses
Long
: 2-word data with sign. This data is handled in 2-word units. One data per four
addresses
float
: Floating data. This data is handled in 2-word units. One data per four addresses
double
: Floating data. This data is handled in 4-word units. One data per eight addresses
7.2.1
Word data [Read-out/Write-in]: Function code [ 03H, 10H ]
Relative
address
Register
No.
Data
type
Parameter
Read-out data/Write-in data setting range
Remarks
4XXXX
0000
40001 int
Damping
1 place after the decimal point, 0.0 to 100.0sec
0002 40003
int
Not
use
0004
40005 int
Flow rate unit
Metric system : 0:L/s, 1:L/min, 2:L/h, 3:L/d,
4:kL/d, 5:ML/d, 6:m
3
/s, 7:m
3
/min, 8:m
3
/h,
9:m
3
/d, 10:km
3
/d, 11:Mm
3
/d, 12:BBL/s,
13:BBL/min, 14:BBL/h, 15:BBL/d,
16:kBBL/d, 17:MBBL/d
English system : 0:gal/s, 1:gal/min, 2:gal/h,
3:gal/d, 4:kgal/d, 5:Mgal/d, 6:ft
3
/s, 7:ft
3
/min,
8:ft
3
/h, 9:ft
3
/d, 10:kft
3
/d, 11:Mft
3
/d, 12:BBL/s,
13:BBL/min, 14:BBL/h, 15:BBL/d,
16:kBBL/d, 17:MBBL/d
0006 40007
int
Range
type
0: Single range, 1: Auto 2 range, 2: Forward-
reverse range, 3: Forward-reverse auto 2 range
0008
40009 double
Full scale 1
Metric system: 64-bit floating data; 0, ±0.3 to
±32m/s as converted to flow rate
English system: 64-bit floating data
Unit: Flow rate
0010
40017 double
Full scale 2
Metric system: 64-bit floating data; 0, ±0.3 to
±32m/s as converted to flow rate
English system: 64-bit floating data
Unit: Flow rate
0018
40025 int
Range hysteresis
2 places after the decimal point, 0.00 to 20.00%
001A
40027 int
Burnout
0: Not use, 1: Hold, 2: Upper, 3: Lower, 4: Zero
001C
40029 int
Burnout timer
Decimal point fixed, 0 to 900sec
001E
40031 int
Output limit low
Decimal point fixed,
20 to 0%
0020
40033 int
Output limit high
Decimal point fixed, 100 to 120%
0022
40035 int
Rate limit timer
Decimal point fixed, 0 to 900sec
0024 40037
double Rate
limit
Metric system: 64-bit floating data; 0 to 5m/s
as converted to flow rate
English system: 64-bit floating data
Unit: Flow rate
002C
40045 double
Low flow rate cut
Metric system: 64-bit floating data; 0 to 5m/s
as converted to flow rate
English system: 64-bit floating data
Unit: Flow rate
0034 40053
double Calibration
zero
Metric system: 64-bit floating data; ±5m/s as
converted to flow rate
English system: 64-bit floating data
Unit: Flow rate
003C
40061 int
Calibration span
2 places after the decimal point, 0.00 to 200.00%
003E
40063 int
Operation mode
0: Normal, 1: High speed response mode
0040 40065
int
Total
unit
*1
Metric system : 0:mL, 1:L, 2:m
3
, 3:km
3
,
4:Mm
3
, 5:mBBL, 6:BBL, 7:kBBL
English system : 0:gal, 1:kgal, 2:ft
3
, 3:kft
3
,
4:Mft
3
, 5:mBBL, 6:BBL, 7:kBBL, 8:ACRf
Bulletin F-107-UXF3-S