Dwyer UXF3 User Manual
Page 65
- 61 -
Table 9-18 < Operation Information >
Item Content
WEDGE SOUND VELOCITY
m/s [ft/s]
WEDGE ANGLE
°
PIPE SOUND VELOCITY
m/s [ft/s]
ANGLE IN PIPE
°
LINING SOUND VELOCITY
m/s [ft/s]
ANGLE IN LINING
°
FLUID SOUND VELOCITY
m/s [ft/s]
ANGLE IN FLUID
°
TOTAL TIME
µs
WINDOW OPEN (Win C)
µs
FORWARD TIME
µs
RESERVE TIME
µs
TRANSIT TIME
ns
DELAY TIME
µs
FLUID SOUND VELOCITY
m/s [ft/s]
ANGLE IN FLUID
°
REINOLDS No. (Re)
K
VELOCITY m/s
[ft/s]
U: SIGNAL POWER (AGC U)
%
* When measurement is normal: 35% or higher
D: SIGNAL POWER (AGC D)
%
* When measurement is normal: 35% or higher
U: SIGNAL PEEK (P/H U)
* When measurement is normal: Stabilizes within the range
from 5528 to 6758.
D: SIGNAL PEEK (P/H D)
* When measurement is normal: Stabilizes within the range
from 5528 to 6758.
U: TRIG. LEVEL (TRG U)
%
D: TRIG. LEVEL (TRG D)
%
See “Displaying data in maintenance mode” in “6. Maintenance and Inspection” of the separate instruction manual,
“FIXED TYPE ULTRASONIC FLOWMETER” Operation Manual, for the operation information.
Bulletin F-107-UXF3-S