Dwyer UXF3 User Manual
Page 62
- 58 -
[Setting]....................... Sends the setting of the selected item (check box set to ON (
;)), reflecting the response
value on the setting.
[READ] ....................... Reads the setting of the selected item (check box set to ON (
;)), reflecting the response
value on the setting.
[Save] .......................... Reflects the setting sent by pressing the [Setting] button on the flow transmitter. Be sure to
press the [Save] button if the setting is changed.
[Check ON/OFF]......... Set the check box to ON to select all the items (to set all the check boxes to ON (
;)). Set
the check box to OFF (
) to release the selection of all the items (to set all the check boxes
to OFF (
)).
Table 9-17 < Detailed Setting >
Item Content
TRANSMIT PATTERN
Select from BURST 1, BURST 2, BURST 3, BURST 4, BURST 5, CHIRP 4, CHIRP 8,
and RESERVE.
TRANSMIT count
In case operation mode is normal:
Select from 8, 16, 32, 64, 128 and 256.
In case operation mode is High Speed:
Select from 4, 8, 16, 32, 64 and 128.
MEASURE METHOD
Select from METHOD 1, METHOD 2 and METHOD 3.
SATURATION
Enter in the range of numeric 0 to 512.
SIGNAL BALANCE
Enter in the range of numeric 0 to 100%.
SIGNAL PEEK
Select from 0.125 V (1024), 0.25 V (2048), 0.375 V (3072), and 0.5 V (4096).
TRIGGER LEVEL
With selection of AUTO/MANUAL, in case of MANUAL, input range of numeric 10.00
to 90.00% at right column.
WINDOW CONTROL
With selection of AUTO/MANUAL, in case of MANUAL, input range of numeric 1 to
16383 in each column of U: OPEN TIME/D: OPEN TIME.
AGC GAIN
With selection of AUTO/MANUAL, in case of MANUAL, input range of numeric 1.28 to
98.56% in each column of U: AGC/D: AGC.
TRANS. WAIT TIME
Enter in the range of numeric 1 to 30 msec.
Bulletin F-107-UXF3-S