The address data indicated below is not stored in the main unit.

Relative

address

Register

No.

Data

type

Parameter

Read-out data/Write-in data setting range

Remarks

0150

40337 int

Set value storage request

Read-out data: 0: Completed, 1: Being stored
Write-in data : 1: Storage

0152 40339

int

Memory

initialize

Read-out data: 0
Write-in data : 100: Initialize

Communication
is disabled for
about 5 seconds
after
initialization.

0154

40341 int

Output setting (current check)

Decimal point fixed,

20 to 120%

0156

40343 int

Total pulse check

Decimal point fixed, 1 to 100PULSE/s

0158

40345 int

DO check

0: ON, 1: OFF

015A 40347

int

DI

check

Read-out data: 10: ON, 20: OFF
Write-in data : 0: Not use, 1: Start

015C

40349 int

Test mode

0: Not use, 1: Set

015E

40351 int

Input data

Decimal point fixed, ±120%

0160

40353 int

Tracking time

Decimal point fixed, 0 to 900sec

0162

40355 int

LCD & LED check

*3

0: Not use, 1: Check

0164 40357

int

Key

test

*3

Read-out data: 10: No key pressing, 20: ESC

key, 40: UP key, 80: ENT key, 100: SET key

Write-in data : 0: Not use, 1: Start

0166 40359

int

Test

cancel

Read-out data: 0: Termination, 1: Being tested
Write-in data : 0: Termination

Cancellation of
testing of
current, total
pulse, DO
check and test
mode

0168

to

0170

40361

to

40369

int

Not use, write-
in inhibited.

*3) No response is made if execution is attempted in a screen other than measure screen.

Bulletin F-107-UXF3-S