GDS-3000 Programming Manual

90

:TRIGger:ALTernate

Set

Query

Description

Sets alternating between source triggers on or off.

Syntax

:TRIGger:ALTernate {OFF | ON |?}

Parameter

OFF

Alternate off

ON

Alternate on

Return parameter

Returns the Alternate trigger status (ON, OFF).

Example

:TRIGger:ALTernate ON

Turns on alternating between source triggers.

:TRIGger:LEVel

Set

Query

Description

Sets or queries the trigger level.

Syntax

:TRIGger:LEVel { | ?}

Parameter

Trigger level value

Return parameter

Returns the trigger level as .

Example

:TRIGger:LEVel 3.30E-3

Sets the trigger level to 330mV/mA.

Note: This command is equal to :TRIGger:HLEVel

:TRIGger:TIMEOut:WHEn

Set

Query

Description

Sets or queries the timeout trigger triggering point.
Note: Only applicable for 500MHz models with the
Timeout trigger APP installed.

Syntax

:TRIGger:TIMEOut:WHEn {HIGH | LOW | EITher | ?}

Related
commands

:TRIGger:TIMEOut:TIMER

Parameter/
Return parameter

HIGH

Sets the triggering point as stays high.

LOW

Sets the triggering point as stays low.