COMMAND DETAILS

41

Return parameter Returns the probe type.

Example

:CHANnel1:PROBe:TYPe VOLTage
Sets the Channel 1 probe type to voltage.

:CHANnel:SCALe

Set

Query

Description

Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Note the probe attenuation factor should be set
before the scale.

Syntax

:CHANnel:SCALe { | ?}

Parameter

Channel

Vertical scale

1/2/3/4 CH1/2/3/4 2e–3 ~ 1e+1 2mV ~ 10V

(Probe x1)

Return parameter Returns the vertical scale in volts or amps.

Example

:CHANnel1:SCAle 2.00E–2
Sets the Channel 1 vertical scale to 20mV/div

Math Commands

:MATH:DISP ............................................................... 42

:MATH:TYPe ............................................................... 42

:MATH:DUAL:SOURce ......................................... 42

:MATH:DUAL:OPERator ............................................ 43

:MATH:DUAL:POSition .............................................. 43

:MATH:DUAL:SCALe .................................................. 44

:MATH:FFT:SOURce .................................................. 44

:MATH:FFT:MAG ........................................................ 45

:MATH:FFT:WINDow ................................................. 45

:MATH:FFT:POSition ................................................. 46

:MATH:FFT:SCALe ...................................................... 46