beautypg.com

Acquire:mode – GW Instek GDS-3000 Series Programming Manual User Manual

Page 34

background image

GDS-3000 Programming Manual

34

Syntax

:ACQuire:AVERage {2 | 4 | 8 | 16 | 32 | 64 | 128 | 256
| ?}

Parameter

2, 4, 8 ,16, 32, 64, 128, 256

Note

Before using this command, select the average
acquisition mode. See the example below.

Example

:ACQuire:MODe AVERage
:ACQuire:AVERage 2
Selects the average acquisition mode, and sets the
average number to 2.

:ACQuire:MODe

Set

Query

Description

Selects or returns the acquisition mode.

Syntax

:ACQuire:MODe {SAMPle | PDETect | HIRes |
AVERage | ?}

Parameter

SAMPle

Sample mode
sampling

PDETect

Peak detect
sampling

HIRes

Hi resolution
sampling

AVERage

Average
sampling mode

Example

:ACQuire:MODe PDETect
Sets the sampling mode to peak detection.

:ACQuire:MEMory?

Query

Description

Returns the data in acquisition memory for the
selected channel as a single string.

Syntax

:ACQuire:MEMory?

Parameter

Channel

1/2/3/4

Channel 1/2/3/4