beautypg.com

Gonogo:script, Template:mode, Template:maximum – GW Instek GDS-3000 Series Programming Manual User Manual

Page 100

background image

GDS-3000 Programming Manual

100

:GONogo:SCRipt

Set

Description

Activates/Deactivates the Go/NoGo APP or
queries its state.

Syntax

:GONogo:SCRipt {OFF | ON | ?}

Parameter/
Return Parameter

ON

Turns Go/NoGo APP on.

OFF

Turns the Go/NoGo APP off.

Example

:GONogo:SCRipt?
>ON
The Go/NoGo script is on.

:TEMPlate:MODe

Set

Query

Description

Sets or returns the Go/NoGo template mode.

Syntax

:TEMPlate:MODe{MAXimum|MINimum|AUTO|?}

Parameter/
Return Parameter

MAXimum

Maximum template

MINimum

Minimum template

AUTO

Auto template

Example

:TEMPlate:MODe

AUTO

Sets the template mode to AUTO.

:TEMPlate:MAXimum

Set

Query

Description

Defines or queries which waveform memory
(REF1 or W1~W20) is set to the maximum
template.

Syntax

:TEMPlate:MAXimum{REF1|W1~W20|?}

Parameter/
Return Parameter

REF1

Reference one

W1~W20 Waveform memory 1 to 20