beautypg.com

Configure gpib interface – GW Instek GDS-3000 Series Programming Manual User Manual

Page 10

background image

GDS-3000 Programming Manual

10

Configure GPIB Interface

To use GPIB, the optional GPIB to USB (GUG-001) adapter must be
used. The GPIB address can be configured for the GUG-001 from the
utility menu. See the GUG-001 user manual for more information.

Configure GPIB

1. Insert the GUG-001 USB 2.0 A-B

type cable into the rear panel USB
device port.

2. Press the Utility key.

Utility

3. Press I/O from the bottom menu.

4. Press USB Device Port from the

side menu and choose Computer.

5. Press GPIB from the side menu.

6. Use the variable knob to set the

GPIB Address from the side menu.

Range

1 ~ 30

GPIB constraints

Maximum 15 devices altogether, 20m cable
length, 2m between each device

Unique address assigned to each device

At least 2/3 of the devices turned On