4-11

Commands

M-4662

1

2

3

4

5

6

7

App

Index

p8

Scaling decimal place (0 to 4)

p9

Unit (up to six alphanumeric characters)

p10

Low-cut function ON/OFF

p11

Low-cut point (0 to 50)

Query

SR[ p1]?

Example

Convert the DC voltage measured on channel
001 to a flow rate using the square root
computation. Set the input range to 6 V, the span
lower limit to 1 V, the span upper limit to 5 V, the
scaling lower limit to 10.0 m

3

/s, and the scaling

upper limit to 100.0 m

3

/s.

SR001,SQRT,6V,1000,5000,100,1000,1,
m3/s

Description • You cannot execute this command while the

MV is measuring, computing, or generating a
report.

• Set parameters p1 and p3 according to the

table in section 4.3.

• For parameters p4 and p5, enter a value using

5 digits or less excluding the decimal point.

• Set all parameters p6, p7, and p8 or omit all

three parameters.

To set a channel to 1-5V DC input

Syntax

SR p1,p2,p3,p4,p5,p6,p7,p8,p9,p10

p1

Measurement channel number

p2

Input type(1-5V)

p3

Measurement range (1-5V)

p4

Span lower limit (800 to 5200)

p5

Span upper limit (800 to 5200)

p6

Scaling lower limit (–30000 to 30000)

p7

Scaling upper limit (–30000 to 30000)

p8

Scaling decimal place (0 to 4)

p9

Unit (up to six alphanumeric characters)

p10

Low-cut function ON/OFF

Query

SR[ p1]?

Example

Set the channel 005 input type to 1-5V, the span
lower limit to 1 V, the span upper limit to 5 V, and
turn the 1-5V low-cut function ON.
SR005,1-5V,1-5V,1000,5000,,,,,ON

Description • You cannot execute this command while the

MV is measuring or computing.

• Set parameter p1 according to the table in

section 4.3.

• For parameters p4 and p5, enter a value using

4 digits or less excluding the decimal point.

• Set all parameters p6, p7, and p8 or omit all

three parameters.

SO

Sets a calculation expression

Syntax

SO p1,p2,p3,p4,p5,p6,p7
p1

Computation channel number

p2

Computation ON/OFF

p3

Calculation expression (up to 120

characters)

p4

Span lower limit (–9999999 to 99999999)

p5

Span upper limit (–9999999 to 99999999)

p6

Span decimal place (0 to 4)

p7

Unit (up to six alphanumeric characters)

Query

SO[ p1]?

Example

Compute the sum of channels 001 and 002
on channel 106. Set the span lower limit to
–10.0000, the span upper limit to 15.0000, and
the unit to V.
SO106,ON,001+002,-100000,150000,4,V

Description • You can execute this command on models

with the /M1 math option.

• You cannot execute this command while the

MV is measuring or computing.

• For details on calculation expressions, see

section 2.2.

• Set parameter p1 according to the table in

section 4.3.

• For parameters p4 and p5, enter a value

using 7 digits or less excluding the decimal
for negative numbers and 8 digits or less for
positive numbers.

• Set all parameters p4, p5, and p6 or omit all

three parameters.

ER

Sets the range of an external

input channel

Syntax

ER p1,p2,p3,p4,p5,p6
p1

External input channel number

p2

External input channel ON/OFF

p3

Span lower limit (–30000 to 30000)

p4

Span upper limit (–30000 to 30000)

p5

Decimal place (0 to 4)

p6

Unit (up to six alphanumeric characters)

Query

ER[ p1]?

Example

Set the span of external input channel 201 to
–150.00 to 150.00.
201,ON,-15000,15000,2

Description You can execute this command on models with

the /MC1 external input channel option.

TJ

Sets memory sampling

Syntax

TJ p1,p2
p1

Measurement, computation, or external

input channel number

p2

Memory sampling ON/OFF

Query

TJ[ p1]?

Example

Save channel 002 to memory.
TJ002,ON

Description You can execute a computation channel (or

make a query) on models with the /M1 math
option. You can specify an external input channel
(or make a query) on models with the /MC1
external input channel option.

4.4 Setting Commands (Setting)