3 binary data output format – Yokogawa Button Operated MV2000 User Manual

Page 145

background image

5-27

M-4662

Responses

1

2

3

4

5

6

7

App

Index

5.3

Binary Data Output Format

This section describes the binary data output format that is disclosed.
For information about other binary data, see section 5.1.
• Instantaneous data (measured/computed/external input) and FIFO data
• Configured channel information data
• Configured alarm information data
• Manual sample file
• Report sample file
• Display data (text)
• Event data (text)
The MV returns measured data and computed data using signed 16-bit integer and
signed 32-bit integer, respectively. These integers can be understood as physical values
by adding a decimal point and unit. The decimal place can be determined using the FE
command.

Examples of Obtaining Physical Values from Binary Data

Binary Value

Decimal Place Code

Physical Value (Measured Value)

10000

0

10000

10000

1

1000.0

10000

2

100.00

10000

3

10.000

10000

4

1.0000

Note

“CRLF” used in this section denotes carriage return line feed.