5-7

M-4662

Responses

1

2

3

4

5

6

7

App

Index

tt

Command name (SR, SA···, XA, XI···)

sss···s

Setting/basic setting data (variable length, one line)

• Example

EA
SR001,VOLT,20mV,0,20
SR002,VOLT,20mV,0,20
···················
EN

Decimal Place and Unit Information

• The MV returns decimal place and unit information in response to an FE command.
• You can use the CB command to specify whether or not the MV will return the data of

measurement channels set to SKIP and computation channels set to OFF.

Syntax

The MV returns the data for each channel in the following syntax.

EACRLF
s_cccuuuuuu,ppCRLF
··················
ENCRLF

s

Data status (N, D, or S)

N:

Normal

D:

Differential input

S:

Skip (When the measurement range is set to SKIP for a

measurement channel or when the channel is turned OFF for a
computation channel)

ccc

Channel number (3 digits)

001

to 048:

Measurement channel

101

to 160:

Computation channel

201

to 440:

External input channel

uuuuuu

Unit information (six characters, left-justified)

mV____:

mV

V_____:

V

^C____:

°C

xxxxxx:

(User-defined character string)

pp Decimal place (00 to 04)
No decimal (00000) for 00.
One digit to the right of the decimal (0000.0) for 01.
Two digits to the right of the decimal (000.00) for 02.
Three digits to the right of the decimal (00.000) for 03.
Four digits to the right of the decimal (0.0000) for 04.
_

Space

Example

EA
N 001mV ,01
N 002mV ,01
EN

5.2 Text Data Output Format