beautypg.com

10 output commands (special response commands), App index – Yokogawa Button Operated MV2000 User Manual

Page 116

background image

4-43

Commands

M-4662

1

2

3

4

5

6

7

App

Index

4.9 Output Commands (RS-

422/485 Commands)

ESC O Opens an instrument

The ASCII code for ESC is 1BH. See appendix 3
for details.

Syntax

ESC O p1
p1

Instrument address (01 to 99)

Example

Open the instrument at address 99, and enable
all commands.
ESC O99

Description • Specifies the address of the instrument to

communicate with.

• You can only open one instrument.
• When you open an instrument with the ESC O

command, another instrument that is currently
open is automatically closed.

• When the MV receives this command

normally, the MV transmits the data “ESC O

.”

• Normally, either CR+LF or LF can be used for

the terminator in communication commands.
However, the terminator for this command
must be set to CR+LF.

ESC C Closes an instrument

The ASCII code for ESC is 1BH. See appendix 3
for details.

Syntax

ESC C p1
p1

Instrument address (01 to 99)

Example

Close the instrument with the address 77.
ESC C77

Description • Releases the connection with the instrument.

• When the MV receives this command

normally, the MV transmits the data “ESC C

.”

• Normally, either CR+LF or LF can be used for

the terminator in communication commands.
However, the terminator for this command
must be set to CR+LF.

4.10 Output Commands (Special

Response Commands)

*I

Transmits instrument

information

Syntax

*I

Description Transmits the maker, model, serial number, and

firmware version in a comma-separated ASCII
text with a terminator at the end.

Example Omega,RD-MV1000,99AA0123,F1.01

4.9 Output Commands (RS-422/485 Commands)/4.10 Output Commands (Special Response Commands)

This manual is related to the following products: