Yokogawa Button Operated MV2000 User Manual
Page 127
5-9
M-4662
Responses
1
2
3
4
5
6
7
App
Index
ddddd
Mantissa (00000 to 99999, 5 digits)
• Eight digits for computed data.
• For abnormal data (data status is E) or data whose mantissa or
exponent exceeds the range (data status is O), the mantissa is set to
99999 (99999999 for computed data).
pp
Exponent (00 to 04)
_
Space
•
Example
EA
DATE 99/02/23
TIME 19:56:32.500
N 001h mV +12345E-03
N 002 mV -67890E-01
S 003
EN
Note
• The MV does not return data for channels that do not exist (not even the channel number).
• For channels set to skip, the values from the alarm status to exponent will be spaces.
Relay and Internal Switch Status
The MV returns the relay status and internal switch status in response to an FD
command.
• Syntax
EACRLF
I01-I06:aaaaaaCRLF
I11-I16:aaaaaaCRLF
I21-I26:aaaaaaCRLF
I31-I36:aaaaaaCRLF
S01-S30:aaa···CRLF
ENCRLF
aaa···
Indicates the relay statuses in ascending order by relay number from the
left.
1:
Relay ON
0:
Relay OFF
-:
Relay not installed
• Example 1
When relays I01 to I04 are ON, and I05 and I06 are not installed (for the RD-MV1000)
EA
I01-I06:1111--
I11-I16:------
I21-I26:------
I31-I36:------
S01-I30:000000000000000000000000000000
EN
5.2 Text Data Output Format