beautypg.com

7 output commands (control), App index – Yokogawa Button Operated MV2000 User Manual

Page 112

background image

4-39

Commands

M-4662

1

2

3

4

5

6

7

App

Index

YE

Activates Basic Setting Mode

(cold reset)

Syntax

YE p1
p1

Setting activation

STORE

Start by saving the basic settings

ABORT

Start without saving basic settings

Example

Start by saving the basic settings.
YESTORE

4.7 Output Commands

(Control)

BO

Sets the byte output order

Syntax

BO p1
p1

Byte order

0

Transmits the data MSB first.

1

Transmits the data LSB first.

Query

BO?

Example

Output the data MSB first.
BO0

Description This command applies to the byte order of

numeric data for BINARY output.

CS

Sets the checksum

Syntax

CS p1
p1

Checksum usage

0

Not calculate (value fixed to zero)

1

Calculate

Query

CS?

Example

Enable (calculate) the checksum.
CS1

Description You can only use this command for serial

communications.

IF

Sets the status filter

Syntax

IF p1,P2
p1

Filter value of status information 1 to 4

(0.0.0.0 to 255.255.255.255)

p2

Filter value of status information 5 to 8

(0.0.0.0 to 255.255.255.255)

Query

IF?

Example

Set the status filter value to 1.0.4.0 and
255.127.63.31.
IF 1.0.4.0,255.127.63.31

Description See chapter 6 for details.

CC

Disconnects the Ethernet

connection

Syntax

CC p1
p1

Disconnection (0)

Example

Disconnect the connection.
CC0

4.6 Basic Setting Commands/4.7 Output Commands (Control)

This manual is related to the following products: