beautypg.com

Effects of operating conditions – Yokogawa Button Operated DX2000 User Manual

Page 340

background image

13-22

IM 04L42B01-01E

Item

Specifications

Measuring accuracy in case of scaling

Accuracy during scaling (digits) = measurement accuracy (digits) × multiplier + 2 digits (rounded up)

* Fractions rounded up

where the multiplier = scaling span (digits)/measuring span (digits).

Example For 1-5 V range (A/D integration time is 16.7 ms or more), measurement span of 1.000 to

5.000 V, and scaling span of 0.000 to 2.000

The measuring accuracy for 5 V input is as follows.

Measuring accuracy (1-5 Vrange) = ±(0.05% × 5 V + 3 digits) = ±(0.0025 V [3 digits] + 3

digits) = ±6 digits

Multiplier = {2000 digits (0.000 to 2.000)}/4000 digits (1.000 to 5.000) = 0.5

Thus, accuracy during scaling = ±(6 × 0.5 + 2) digits = 5 digits (rounded up)

Reference junction compensation accuracy

When measuring temperature greater than or equal to 0 °C and when input terminal temperature is balanced

Type R, S, W, WRe:

±1.0°C

Type K, J, E, T, N, L, and U: ±0.5°C. Type B: Internal reference compensation is fixed to 0°C

Maximum input voltage ±60 VDC (continuous)

Input resistance

200 mV range or less and TC: 10 MΩ or more

2 V range or higher: Approx. 1 MΩ

Input source resistance

Volt, TC

2 kΩ or less

RTD input

10 Ω or less per wire (The resistance of all three wires must be equal).

Bias current

10 nA or less (except when burnout detection function is enabled)

Maximum common mode noise voltage

250 VACrms (50 Hz/60 Hz)

Maximum noise voltage between channels

250 VACrms (50 Hz/60 Hz)

Interference across channels

120 dB (when the input source resistance is 500 Ω and the input to other channels is 60 VDC)

Common mode rejection ratio

When the A/D integration time is 20 ms

120 dB (50 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)

When the A/D integration time is 16.7 ms

120 dB (60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)

When the A/D integration time is 1.67 ms

80 dB or higher (50/60 Hz ± 0.1%, 500 Ω unbalanced, between the minus terminal and ground)

Normal mode rejection ratio

When the A/D integration time is 20 ms

40 dB or more (50 Hz ± 0.1%)

When the A/D integration time is 16.7 ms

40 dB or more (60 Hz ± 0.1%)

When the A/D integration time is 1.67 ms

Not reject 50/60 Hz

Effects of Operating Conditions

Item

Specifications

Ambient temperature (applies when the A/D integration time is 16.7 ms or greater, with temperature variation of 10°C)

DC voltage, TC range ±(0.1% of rdg + 0.05% of range) or less

* Excluding the error of reference junction compensation

RTD range

±(0.1% of rdg + 2 digits) or less

Power supply fluctuation With variation within 90 to 132 V and 180 to 250 VAC (50/60 Hz):

Accuracy specifications are satisfied.

With variation of ±2 Hz from rated power frequency (power supply voltage 100 VAC):

Accuracy specifications are satisfied.

Magnetic field

AC (50/60 Hz) and DC 400 A/m fields:±(0.1% of rdg + 10 digits) or less

Input source resistance

DC voltage range

With variation of +1 kΩ:

200 mV range or less: ±10 µV or less

2 V range or higher: ±0.15% of rdg or less

TC range

With variation of +1 kΩ: ±10 µV or less

RTD range (Pt100)

With variation of 10 Ω per wire (resistance of all three wires must be equal): ±(0.1% of rdg + 1 digits)

or less

With maximum difference of 40 mΩ between wires: Approx. 0.1 °C

Effects of vibration

Effects from a sinusoidal vibration along all three axis at a frequency between 10 to 60 Hz and an

acceleration of 0.2 m/s

2

: ±(0.1% of rdg + 1 digit) or less

13.6 General Specifications

This manual is related to the following products: