App index – Yokogawa Button Operated DX2000 User Manual
Page 243
6-23
IM 04L42B01-01E
Saving and Loading Data
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
Explanation
• Setup Data File
On DXs without the /AS1 advanced security option
• The setup data file extension is .PDL.
• The maximum setup data file size is approximately 250 KB.
• The following settings are also saved.
• Current monitor display conditions
• Default display registration data
• Favorite key registration data
On DXs with the /AS1 advanced security option
• The setup data file extension is .PEL.
• The maximum setup data file size is approximately 250 KB.
• The following settings are also saved.
• Current monitor display conditions
• Default display registration data
• Favorite key registration data
• Loading Setup Data
On DXs without the /AS1 advanced security option
• Only the setup data of the setting mode is loaded in the setting mode. However,
settings that contradict the setup data of the basic setting mode are not loaded.
• The monitor display conditions, default display, and favorite keys are also loaded.
• If the contents of the loaded setup data is invalid, check the error log (see section
4.9).
• Operations through keys, communications, and remote control input are not
executed while the setup data is being loaded.
On DXs with the /AS1 advanced security option
• If the setup data is changed by loading new setup data, the original setup file
is automatically saved to CF card. If no CF card is inserted, an error message
appears and the operation cannot be completed.
• All settings, Login info only, and Other settings
The “Load settings”
option
Description
All settings
The DX loads all setting mode and basic setting mode settings.
However, the login password is not loaded. The password is set
to the default password.
Login info only
The DX only loads the login settings. However, the login
password is not loaded. The password is set to the default
password.
Other settings
The DX loads all setting except for the login settings.
• In setting mode, only the setting mode settings are loaded. However, settings that
contradict the basic setting mode settings are not loaded.
• In any case, the option to “Load settings” also loads monitor display conditions, the
default display, and favorite keys.
• If the loaded setup data is not applied, check the error log (see section 4.9).
Note
• While loading the setup data, key operations, operations via communications, and
operations via remote input are not available.
• While loading the setup data, actions of the Event Action function are disabled. Events
occurring while loading the setup data are ignored.
6.9 Saving/Loading the Setup Data