Computation and Report Functions (/M1 and /PM1 Options)

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

• Span Lower, Span Upper

Set the measurement range.
Selectable range of values: –9999999 to 99999999
Selectable decimal places: X.XXXX, XX.XXX, XXX.XX, XXXX.X

• Unit

Set the unit of the computed value (Up to 6 characters,

A a # 1

).

• Alarm

The available alarm types are high limit alarm, low limit alarm, delay high limit alarm,

and delay low limit alarm.

The range of alarm values is as follows:

Type

Value

H, L, T, t

Within –9999999 to 99999999 excluding the decimal point

For details on setting alarms, see section 3.7.

* If the Math On/Off or calculation expression is changed, the alarms for that channel are turned

Off.

• Alarm delay > Time

Set the alarm delay time using an integer in the range of 1 to 3600 s.

• Tag > Comment

Sets the tag comment. On a DX with a release number of 3 or later, you can enter up

to 32 characters. On a DX with a release number of 2 or earlier, you can enter up to
16 characters. The characters that can be entered are:

A a # 1

.

• Tag > No. (Release number 3 or later)

This setting only appears when you have enabled the use of tag numbers.
Sets the tag number. (Up to 16 characters:

A a # 1

)

• Constant
• Number of constant

Select the constant (K01 to K60) to set.

• Value

The selectable range is as follows:
–9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
The number of significant digits of a constant is five. When specifying the

constant using exponential notation, set the mantissa less than or equal to 5
digits and the exponent less than or equal to 2 digits.

• TLOG
• Timer type

You can set the timer type to

Timer or Match T (match time timer).

• Timer No.

Select the timer number to use.
For details on setting the timer, see section 7.1.
For match time timer setting instructions, see section 7.1.

• Sum scale

Set the sum scale to

/s to /h to match the unit of the measured value.

Example: If the unit of the measured value is “m

3

/min,” select

/min.

Off:

Sums as-is the measured data per scan interval.

• Reset

To reset the TLOG computed value at each interval, select

On.

9.1 Setting the Expression, Measurement Range, Alarm, Tag, and Data Storage on Computation Channels