Yokogawa Button Operated DX2000 User Manual
Page 233
6-13
IM 04L42B01-01E
Saving and Loading Data
1
2
3
4
5
6
7
8
9
10
11
12
13
App
Index
• Changing Settings and Performing File Operations during Recording
(Memory sampling)
If you are using the multi batch function (/BT2 option), if even one batch group is being
recorded, the DX is recording (memory sampling).
On DXs without the /AS1 advanced security option
You can change all settings except for the ones listed below.
• Basic settings
• Input range
• Memory sampling on/off
• Computation channel calculation expressions and constants
• TLOG
• Trend interval
• File save interval
• Timer and match time timer
On DXs With the /AS1 Advanced Security Option
You can change the following settings and perform the following file operations. The
administrator can perform all operations. Users can only perform operations that have
been permitted. The setting menu that appears varies depending on the operations
that can be performed.
Setting Changes
• Date and time settings
• Alarm settings
• Alarm delay time settings
• Calibration settings
• Destination directory settings
• Administrator settings*
• User settings*
* See the Advanced Security Function (/AS1) User’s Manual, IM 04L41B01-05EN.
File Operations
• Loading display data files
• Loading event data files
• Listing the files on the external storage medium
6.4 Starting/Stopping the Recording and Saving the Measured data