beautypg.com

6 general specifications, General specifications -19, App index – Yokogawa Button Operated DX2000 User Manual

Page 337: Construction, Normal operating conditions, Power supply, Isolation

background image

13-19

IM 04L42B01-01E

Specifications

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

13.6 General Specifications

Construction

Item

Specifications

Mounting

Flush panel mounting (on a vertical plane) (excluding the desktop type)

Mounting angle

Inclined backward up to 30 degrees from a horizontal plane.

Allowable panel thickness

2 to 26 mm

Material

Case:

Metal plate

Bezel and display cover: Polycarbonate

Color

Case:

Grayish blue green (Munsell 2.0B5.0/1.7 or equivalent)

Bezel:

Charcoal gray light (Munsell 10B3.6/0.3 or equivalent)

Front panel

Water and dust proof: Complies with IEC529-IP65 and NEMA No.250 TYPE 4 (except External

Icing Test) (Style number 2 or later), except for side-by-side mounting

External dimensions

288(W) × 288(H) × 226(D) mm (D: depth from the panel mounting plane)

Weight

DX2004 and DX2010: Approx. 6 kg. DX2008 and DX2020: Approx. 6.3 kg. DX2030: Approx. 6.9

kg, DX2040 and DX2048: Approx. 7.3 kg excluding options

Normal Operating Conditions

Item

Specifications

Supply voltage

90 to 132, 180 to 250 VAC

Power supply frequency

50 Hz ± 2%, 60 Hz ± 2%

Ambient temperature

0 to 50°C

Ambient humidity

20 to 80%RH (at 5 to 40°C), 10 to 50% (at 40 to 50°C)

Vibration

10 to 60 Hz, 0.2 m/s

2

Shock

Not allowed

Magnetic field

400 A/m or less (DC and 50/60 Hz)

Noise

Normal mode (50/60 Hz)

DC voltage

The peak value including the signal must be less than 1.2 times the measuring range.

Thermocouple

The peak value including the signal must be less than 1.2 times the measuring thermal electromotive force.

RTD

50 mV or less

Common mode noise

250 VACrms or less for all ranges (50/60 Hz)

Maximum noise voltage between channels

250 VACrms (50/60 Hz) or less

Mounting position

Can be inclined up to 30 degrees backward. Left and right horizontal.

Warm-up time

At least 30 minutes after power on

Installation location

Indoors

Operating altitude

2000 m or less

Power Supply

Item

Specifications

Rated supply voltage

100 to 240 VAC

Allowable power supply voltage range

90 to 132, 180 to 264 VAC

Rated power supply frequency

50 Hz to 60 Hz

Power consumption

Supply voltage

LCD backlight off

Normal

Maximum

100 VAC

28 VA

42 VA

74 VA

240 VAC

38 VA

54 VA

100 VA

Allowable interruption time

Less than 1 cycle of the power supply frequency

Isolation

Item

Specifications

Insulation resistance Between the Ethernet, RS-422/485, and insulation terminals and earth: 20 MΩ or greater at 500 VDC
Withstand voltage

Between the power terminal and earth:

2300 VAC at 50/60 Hz for one minute

Between the contact output terminal and earth:

1600 VAC at 50/60 Hz for one minute

Between the measurement input terminal and earth: 1500 VAC at 50/60 Hz for one minute
Mutually between measurement input terminals:

1000 VAC (50/60 Hz) for one minute (excluding the

RTD input terminal of DX2010, DX2020, DX2030,

DX2040, and DX2048)

Between the remote input terminal and earth:

1000 VDC for one minute

Between the pulse input terminal and earth:

1000 VDC for one minute

Ground

Grounding resistance: 100 Ω or less

This manual is related to the following products: