beautypg.com

Chapter 11 troubleshooting, 1 a list of messages, A list of messages -1 – Yokogawa Button Operated DX2000 User Manual

Page 294: App index

background image

11-1

IM 04L42B01-01E

Troubleshooting

1

2

3

4

5

6

7

8

9

10

11

12

13

App

Index

11.1 A List of Messages

There are cases in which error codes and messages are displayed on the screen during
operation. A list of the possible error codes and messages are given in the table below.
Communication error codes and messages are also listed.
Error responses to communication commands are output in English.

Errors Related to Parameter Settings

Setting Errors

Code

Message

Explanation/Countermeasures/Ref. section

1

System error.

Contact your nearest YOKOGAWA dealer.

2

Incorrect date or time setting.

Enter a correct value.

3

A disabled channel is selected.

Specify a channel that is not set to Skip or Off.
Specify a channel that is installed.

4

Incorrect function parameter.

See chapter 3 the communication manual.

5

The input numerical value exceeds the set range.

Enter a proper value.

6

Incorrect input character string.

Enter a proper character string.

7

Too many characters.

Enter the correct number of characters.

8

Incorrect input mode.

Specify a correct mode.
See section 3.3.

9

Incorrect input range code.

Specify a correct range code.
See section 3.3.

11

Range settings are not same within the selected channels.

Specify channels with the same range setting.
See section 3.9.

12

A disabled batch group is selected.

Set the last batch group greater than the first batch

group.
See section 2.2 in IM04L41B01-03E.

13

Cannot set a trip line for a display group that is OFF.

Check the display group settings.
See section 5.1 and section 2.2 in IM04L41B01-03E.

21

Cannot set an alarm for a skipped channel.

Cannot be specified on channels set to Skip.
See section 3.7.

22

The upper and lower span limits are equal.

Cannot be set to the same value.
See section 3.3.

23

The upper and lower scale limits are equal.

Cannot be set to the same value.
See section 3.3.

24

The lower limit of the span band is greater than the upper limit. Set the lower limit less than the upper limit.

See section 3.3.

25

The lower limit of the scale band is greater than the upper limit. Set the lower limit less than the upper limit.

See section 3.3.

30

The partial boundary value exceeds the range of the span.

Set the boundary value in the range of “the minimum

span value + 1 digit” to “the maximum span value – 1

digit.”
See section 5.9.

31

Partial-expansion display is set ON for a SKIPPED channel.

Cannot be specified on channels set to Skip.
See sections 3.3 and 5.9

35

The upper and lower limits of the display band are equal.

Set the upper limit greater than the lower limit + 5.
See section 5.6.

36

The lower limit of the display band is greater than the upper

limit.

Set the upper limit greater than the lower limit + 5.
See section 5.6.

37

The display band is narrower than 4% of the entire display.

Set the upper limit greater than the lower limit + 5.
See section 5.6.

40

Incorrect group set character string.

Check the syntax.
See section 5.1.

41

There is no specified input channel.

Specify a channel that is installed.
Operation Guide and section 5.1.

42

Exceeded the number of channels which can be set.

Up to 10 channels per group.
See section 5.1.

Chapter 11 Troubleshooting

This manual is related to the following products: