Item
Specifications
Electrical and mechanical specifications
Conforms to IEEE 802.3 (Ethernet frames conform to the DIX specification).
Medium
Ethernet (10BASE-T)
Implemented protocols
TCP, IP, UDP, ICMP, ARP, DHCP, HTTP, FTP, SMTP, SNTP, Modbus, and DX-dedicated protocols
E-mail client
Automatically send e-mail at specified times.
FTP client
Automatically transfer data files to the FTP server.
Applicable files: Display data, event data, screen image data, and report data
FTP Server
Transfer files, delete files, manipulate directories, and output file lists of the DX.
Web server
Shows the DX display on a Web browser.
SNTP client
Inquires the time to the SNTP server and sets the DX.
Cannot be used after hour 0 on January 1st, 2036.
SNTP server
Outputs the DX time.
Time resolution: 5 ms
Cannot be used after hour 0 on January 1st, 2036.
DHCP client
Automatically obtain the network address settings from the DHCP server.
Modbus client
Reads data from another device and writes to the registers.
Modbus server
Loads measurement and computation channel data
Loads and writes external input channel data
Loads and writes communcation input data
Some control commands such as memory start
Modbus client register access limitations
Setting/Measurement server
Operate, set, and output data of the DX using a dedicated protocol.
Maintenance/test server
Outputs connection information and network information.
Instrument information server
Outputs information (serial number, model name, etc.) of the connected DX.
EtherNet/IP server
Can join an EtherNet/IP network as an adapter (server)
Loads measurement and computation channel data
Loads and writes external input channel data
Loads and writes communcation input data
Batch Function
Item
Specifications
Function
Data management using batch names. Enter text fields and batch comments in the data file.
Batch name
Added to the file name of the display data and event data.
Structure
Batch number (up to 32 characters) + lot number (up to 8 digits)
Text field
Adds text to the display data and event data. There are 24 available text fields for release numbers
3 and later. There are 8 available text fields for release numbers 2 and earlier. Up to 20 title
characters and 30 other characters can be entered per field.
Batch comment
Adds text to the display data and event data.
13.4 Other Standard Functions