Execution Errors

Code

Message

Explanation/Countermeasures/Ref. section

150

This action is not possible because sampling is in progress.

Stop the memory sampling and then execute.
See section 6.4.

151

This action is not possible during sampling or calculating.

Stop the memory sampling and then execute.
See sections 6.4 and 9.4.

152

This action is not possible because saving is in progress.

Wait until the saving is complete.

153

This action is not possible because formatting is in progress.

Wait until the formatting is complete.

154

Message not accepte because message limit was reached.

The limit is 50 messages.
See section 5.4.

155

The message is not written while sampling is stopped.

Start the memory sampling and then execute.
See section 6.4.

156

There are no channels to be saved to the memory.

Set the channels to be saved.
See sections 6.1 and 9.1.

157

This function is not possible at this time.

Check the DX status.

158

Exceeds time deviation setting.

When synchronizing the clock through remote control.

When you specified time that exceeds the time deviation

limit on DXs with Advanced security function (/AS1

option). (Firmware version number 4.11 and later)
See section 2.3.

159

It is outside the postscript message write-in range.

Add message can be written to the past section of the

data being memory sampled.
See section 5.4.

162

Data storage is already started.

Memory sampling is in progress on the batch group.
See section 3.2 in IM04L41B01-03E.

163

Data storage is already stopped.

Memory sampling on the batch group has already been

stopped.
See IM04L41B01-03E.

164

This action is not possible because there is a bar code data

error.

Check the text that the barcode reader scanned.
See section 2.11.

170

End process can’t proceed, because setting file is not saved to

Media.

When you switch from setting mode to operation mode

or basic setting mode, the DX automatically saves the

changed settings to the CF card. This message appears

when the DX cannot save the changed settings. Check

the CF card.
See section 1.5 in the advanced security manual.

173

Data save is not possible because of insufficient media

capacity.

Change the external storage medium.

174

Cannot execute because of failure to save unsaved setting file. The DX cannot save an unsaved setting file to the CF

card. Check the CF card.
See the advanced security manual.

175

The calibration due date has not been set properly.

Check the year, month, and day of the calibration due

date.
See section 3.13

700

The specified command does not exist.

The command specified from the custom display is not

registered. Register commands whose command type

is “E-M,” and specify the numbers of those commands

in the custom display.
See section 1.10 in the communication manual.
See section 2.32 in IM04L41B01-04E.

701

Saving data.

The previous transmission is not finished.
See section 1.10 in the communication manual.

702

Failed to save data.

Check for communication errors or Modbus server

errors.
See section 1.10 in the communication manual.

760

Cannot find KDC server.

The KDC server cannot be found in the same domain.
See section 1.14 in the communication manual.

761

KDC server connection error.

An error occurred while the DX was connecting to the

KDC server.
See section 1.14 in the communication manual.

764

Not supported by this machine.

Not supported by the DX.
See section 1.14 in the communication manual.