beautypg.com

Yokogawa digitalYEWFLO (DY) User Manual

Page 83

background image

<9. OPERATION VIA HART CONFIGURATION TOOL (HART 7)>

9-11

IM 01F06A00-01EN

• Simulation Setting and Correlation of Output Value

: Without option code /MV

Simulation

Setting value

Output value

Flow rate

Total fl ow rateProcess temperatureProcess densityContact outputPV % RangeLoop Current

Flow rate

Yes

Yes

No

No

Yes

Yes

Yes

Total fl ow rate

No

Yes

No

No

No

No

No

Process

temperature

No

No

Yes

No

No

No

No

Process density

No

No

No

Yes

No

No

No

PV % Range

No

No

No

No

No

Yes

Yes

Loop Current

No

No

No

No

No

No

Yes

Yes : Simulation value or calculation result of Simulation.
No : Actual process value or parameter setting value.

: With option code /MV
: Function == “Monitor only”
: Function == “Saturated Steam” or “Superheat

Steam” or “LIQUID:Mass”

: Function == “Gas:STD/Normal”
: Function == “Not use”

Simulation

Setting value

Output value

Flow rate

Total fl ow rate

Process temperature

Process density

Contact output

Flow rate

Yes

Yes

No

No

Yes

Total fl ow rate

No

Yes

No

No

No

Process

temperature


No


No

Yes


Yes


No


Yes


Yes


No


Yes

Process density


No


No

No

Yes


No


Yes


Yes


Yes

PV % Range

No

No

No

No

No

Loop Current

No

No

No

No

No

Simulation

Setting value

Output value

PV= Instaneous fl ow rate

PV= Process temperature

PV % Range

Loop Current

PV % Range

Loop Current

Flow rate

Yes

Yes


No


No


Not available


Not available

Total fl ow rate

No

No


No


No


Not available


Not available

Process

temperature


No


No


Yes


Yes


Yes


Yes


Not available


Not available

Process density


No


No


No


No


Yes


Yes


Not available


Not available

PV % Range

Yes

Yes


Yes


Yes


Not available


Not available

Loop Current

No

Yes


No


Yes


Not available


Not available