1) item a : indication, 2) item b : easy setting – Yokogawa digitalYEWFLO (DY) User Manual
Page 38

<6. PARAMETERS>
6-2
IM 01F06A00-01EN
(1) Item A : Indication
These items are for the indication of fl owrate and total.
Item
Name
R / W
Data Range
Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
A00 DISPLAY
Menu A (Display)
A10 FLOW RATE (%)
R
0.0 to 110.0
%
1
Flow rate
A20 FLOW RATE
R
0.0 to 65535
FU+C40
0 to 5
Flow rate (in engineering unit)
A30 TOTAL
R
0 to 999999
(*2)
FU
0 to 5
Totalized value
(Indicate only for /MV and B50 : TEMP)
A40 TEMP (%)
(*1)
R
0.0 to 110.0
%
1
Temperature Values (%)
(Indicate only for /MV)
A41 TEMPERATURE
(*1)
R
–999.9 to 999.9
D20
1
Temperature Values
A60 SELF CHECK
R
GOOD
ERROR
Self-diagnostic message
FU : Flow unit
/MV : Multi-Variable (Build-in Temperature Sensor) Type
(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : There will be linked to the value of B45, it is displayed “E” shows multipiler 10.
(2) Item B : Easy Setting
These items are for the principal items to operate digitalYEWFLO.
A value in “( )” is the data corresponding to the indicator.
Item
Name
R / W
Data Range
Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
B00 EASY SETUP
Menu B
B10 FLOW SPAN
W
0.00001 to 32000
FU + C40
0 to 5 Flow Span
10
D
L
B15 DAMPING
W
0 to 99
sec
0
Damping Time
4
D
L
B20 CONTACT OUT
W
OFF
SCALED PULSE
UNSCALED PULSE
FREQUENCY
ALARM
FLOW SW(LOW:ON)
FLOW SW(LOW:OFF)
(0)
(1)
(2)
(3)
(4)
(5)
(6)
Contact Output Type
(0)
D
L
(Indicate and Set only for B20: SCALED PULSE, UNSCALED PULSE)
B21 PULSE RATE
W
0.00001 to 32000
FU / P
0 to 5 Pulse Output Rate
1.0
(*3)
D
L
(Indicate and Set only for B20: FREQUENCY)
B22 FREQ AT 100%
W
0 to 10000
PPS
0
Pulse Output Rate at sec /100% 1000
D
L
(Indicate and Set only for B20: FLOW SW (ON), FLOW SW (OFF))
B23 SET LEVEL
W
0.00001 to 32000
FU +C40
0 to 5 Flow Switch (Actual Flow rate)
0
D
L
B30 UPPER DISP
W FLOW RATE (%)
FLOW RATE
TEMP (%)
(0)
(1)
(2)
Selection of Upper Display
(only for /MV)
(0)
(*3)
D
L
B31 LOWER DISP
W BLANK
TOTAL
TEMP
(0)
(1)
(2)
Selection of Lower Display
(only for /MV)
(0)
D
L
B40 TOTAL START
W STOP
START
(0)
(1)
Start / Stop of Totalizer
(0)
D
L
B45 TOTAL RATE
W
0.00001 to 32000
(0)
FU / P
0 to 5 Total Rate
1.0
(*3)
D
L
B47 TOTAL RESET
W
NOT EXECUTE
EXECUTE
(0)
(1)
Totalizer Reset
(0)
D
L
(Indicate and Set only for /MV)
B50 A / OUT SELECT
W
FLOW
TEMP
(0)
(1)
Selection of Analog Output
(0)
D
L
(Indicate and Set only for /MV and B50: TEMP)
B51 TEMP 0%
W
-999.9 to 999.9
D20
1
Set Temperature Value at 0%
-40
D
L
B52 TEMP 100%
W
-999.9 to 999.9
D20
1
Set Temperature Value at 100% 250
(*2)
D
L
B60 SELF CHECK
R
GOOD
ERROR
Self-diagnostic message
FU : Flow unit
(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : If 7.00 or less version that can be checked in K50, Initial Value is 260.