7) item h : adjust, 8) item j : test – Yokogawa digitalYEWFLO (DY) User Manual
Page 45

<6. PARAMETERS>
6-9
IM 01F06A00-01EN
(7) Item H : Adjust.
These items are for setting of adjustment.
A value in “( )” is the data corresponding to the indicator.
Item
Name
R / W
Data Range
Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
H00 ADJUST
Menu H (Adjust)
H10 TRIM 4mA
W -1.00 to 1.00
%
2
Trimming 4mA
0.0
D
H11 TRIM 20mA
W -1.00 to 1.00
%
2
Trimming 20mA
0.0
D
H20 USER ADJUST
W 0.00001 to 32000
0 to 5
User Adjust
1.0
D
H25 REYNOLDS ADJ
W NOT ACTIVE
(0)
ACTIVE (1)
Reynolds Coeffi cient
(0)
D
(Indicate and Set only for H25: ACTIVE)
H26 DENSITY f
W 0.00001 to 32000
D25
0 to 5
Density at operating condition
1024
D
H27 VISCOSITY
W 0.00001 to 32000
mPa.s
0 to 5
Viscosity factor
1.0
D
H30 EXPANSION FA
W NOT ACTIVE
(0)
ACTIVE (1)
Expansion correction for
compressible Gas
(0)
D
H40 FLOW ADJUST
W NOT ACTIVE
(0)
ACTIVE (1)
Instrumental Error Adjust
(0)
D
(Indicator and Set only for H40: ACTIVE)
H41 FREQUENCY 1
W 0 to 32000
Hz
0 to 5
First break-point frequency
(f1)
0.0
D
H42 DATA 1
W -50.00 to 50.00
%
2
First correcting value
(d1)
0.0
D
H43 FREQUENCY 2
W 0 to 32000
Hz
0 to 5
Second break-point frequency (f2)
0.0
D
H44 DATA 2
W -50.00 to 50.00
%
2
Second correcting value
(d2)
0.0
D
H45 FREQUENCY 3
W 0 to 32000
Hz
0 to 5
Third break-point frequency
(f3)
0.0
D
H46 DATA 3
W -50.00 to 50.00
%
2
Third correcting value
(d3)
0.0
D
H47 FREQUENCY 4
W 0 to 32000
Hz
0 to 5
Fourth break-point frequency
(f4)
0.0
D
H48 DATA 4
W -50.00 to 50.00
%
2
Fourth correcting value
(d4)
0.0
D
H49 FREQUENCY 5
W 0 to 32000
Hz
0 to 5
Fifth break-point frequency
(f5)
0.0
D
H50 DATA 5
W -50.00 to 50.00
%
2
Fifth correcting value
(d5)
0.0
D
H60 SELF CHECK
R
GOOD
ERROR
Self-diagnostic message
(*1) : Available for 3.10 or greater version that can be checked in K50.
(8) Item J : Test
These items are for test of output.
A value in “( )” is the data corresponding to the indicator.
Item
Name
R / W
Data Range
Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
J00 TEST
Menu J (Test)
J10 OUT ANALOG
W 0.0 to 110.0
%
1
Current Output
0.0
D
J20 OUT PULSE
W 0 to 10000
PPS
0
Pulse Output
0
D
J30 OUT STATUS
W OFF
(0)
ON (1)
Status Output
(0)
D
J40
(*2)
RELEASE TIME
W 10min
(0)
30min (1)
60min (2)
3h
(3)
6h
(4)
12h (5)
Test auto release time
0
D
J60 SELF CHECK
R
GOOD
ERROR
Self-diagnostic message
(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : Available for 7.00 or greater version that can be checked in K50.