(5) Item E : Detector Setup
These items are for detector that has been already set before.
A value in “( )” is the data corresponding to the indicator.
Item
Name
R / W
Data Range
Unit
Decimal
Point
Remark
Initial
Value
Disp. U / D
(*1)
E00 METER SETUP
Menu E (Detector setup)
E10 NOMINAL SIZE
W 15mm
25mm
40mm
50mm
80mm
100mm
150mm
200mm
250mm
300mm
400mm
(0)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
Selection of Nominal Size
(1)
(*2)
D
L
E20 BODY TYPE
W Standard
High Pressure
Low Flow Unit (1)
Low Flow Unit (2)
(0)
(1)
(2)
(3)
Selection of Body Type
(0)
D
L
E30 SENSOR TYPE
W Standard
High Temperature
Low Temperature
(0)
(1)
(2)
Selection of Sensor Type
(0)
D
L
E40 K-FACT UNIT
W P/l
P/USgal
P/UKgal
(0)
(1)
(2)
Selection of K-factor Unit
(0)
D
L
E41 K-FACTOR
W 0.00001 to 32000
E40
0 to 5
K-factor value of 15 deg C
68.6
D
E50 DETECTOR No.
W 16 characters
Detector number
E60 SELF CHECK
R
GOOD
ERROR
Self-diagnostic message
FU : Flow unit
(*1) : Available for 3.10 or greater version that can be checked in K50.
(*2) : If specifi ed when ordering, it is set to the specifi ed contents.