background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

88

9650-001356-01 Rev. D

Removing the Defibrillator Boards and Capacitor Bank
Modules

Tools Required

#1 Phillips screwdriver with torque capability

#2 Phillips screwdriver with torque capability

26 mm open end wrench

Large diagonal cutters

5k Ohm, 25 watt, resistor

Setup

This procedure should be conducted on a static mat.

Wear a wrist strap to ground during the procedure.

Complete these procedures in the following order before performing this procedure:

1.

“Removing the Handle” on page 63

2.

“Removing the Printer Module” on page 64

3.

“Disassembling the Main Unit” on page 66

To reinstall the defibrillator boards and/or capacitor bank modules, reverse
the steps.

Defibrillator Boards and Capacitor Bank Disassembly

1

Lay back end of separated
unit down with opened side
facing up.

2

Pull the black snap rivet
straight out.