Defibrillator pacer module – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 130

C
HAPTER
5
F
UNCTIONAL
D
ESCRIPTION
124
9650-001356-01 Rev. D
End tidal CO
2
(CO
2
)
X Series supports an Oridion
®
Microstream
®
CO
2
module. The CO
2
module mounts to the
Monitor board and communicates to the board through a cable connected to J50. Serial 
communications from the monitor board is implemented via a UART on the PP. The CO
2
module obtains gas samples from the sampling lines through tubing and a 3-wire cable 
connected to the left side panel.
Thermometry
X Series supports two simultaneous temperature measurements. YSI400/YSI700 series probes 
are used to make temperature measurements. Data is communicated to the PP from the 
Temperature ADC via and isolated SPI bus through the FPGA. The Temperature probes 
connect to the monitor board through the IBP/Temp wire harness which connects to J33 of the 
Monitor board. 
Power Circuitry, Battery Charger, DC-DC converters
Power circuitry for all subsystems but the defibrillator-pacer module reside on the Monitor 
Board. Auxiliary Power and Battery Power are automatically switched to provide constant 
current, the unit draws current from auxiliary power when present. There is a smart battery 
charging circuit that will charge the battery when Auxiliary power is present. Switching mode 
DC-DC converters are utilized to make internal 6V, 5V, 3.3V, 1.8V and processor core voltages. 
Regulators are used to provide regulated supply voltages with low dropout for low current 
consumption circuits. Each of the isolated modules, Temp/IBP, ECG, SpO2 have independent 
power supplies. Isolated power supplies are designed for isolating Patient Applied Parts. The 
MP via the FPGA GPIO has the ability to shut down voltage rails or power to individual patient 
parameters. Power from the battery connects to the Monitor board through J29. Auxiliary 
power connects to the Monitor board through J31.
USB Host Port
X Series Monitor Board incorporates one USB Host ver 1.1 port. The port uses the USB 
controller built into the PXA270 Main processor. Full speed of 12 Mbps and Low speed 
1.5Mbps are supported. The Host port supports a USB flash memory stick for data export. The 
USB host connector is in the right side panel and connects to the Monitor board through J44.
Defibrillator Pacer Module
The Defibrillator/Pacer Module consists of two PCB assemblies connected by a flex cable. This 
Defibrillator/Pacer module relies on proven core defibrillation/pacing technology for improved 
power conversion efficiency, reduced EMI, and improved stability in volume production. This 
module contains power supplies, charging and discharge circuitry, pacing circuitry, and the 
Pads/Paddles ECG preamplifier and impedance circuitry. Communication with the 
Defibrillator/Pacer module is a function of the PP. The ECG-Charger board contains the 
charger circuit, ECG acquisition circuitry, and protection components. The 
Processor-Bridge-Pacer board contains the micro controller, CPLD, Pacer, and defibrillation 
shock delivery circuit.