ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 56
C
HAPTER
2
T
ROUBLESHOOTING
50
9650-001356-01 Rev. D
Error Message
Explanation
Technical Action
ALARM SILENCE SHORT
Alarm Silence/Reset key is
stuck.
Replace the front panel.
BLUETOOTH PHONE
UNAVAILABLE
The X Series unit was
unable to establish a Wide
Area Network connection.
None of the Bluetooth
phones paired with the
device were detected.
To correct the problem,
verify the following:
•
The Bluetooth device is
paired with the X Series
unit.
•
The phone is on and
near enough to the
device to permit
communication.
•
The phone is configured
to allow data connection
requests (if the phone
has such a configuration
setting).
CELLULAR CONNECT
FAILURE
The X Series unit was
unable to establish a Wide
Area Network connection. A
cellular modem or Bluetooth
cellular phone connection
was attempted, but failed.
To correct the problem,
verify the following:
•
The cellular provider
information, particularly
the call number, is
correctly configured.
•
The cellular device has
an active plan.
•
There is adequate
cellular network signal
coverage in the area.
•
(For phones) The phone
is configured to allow
data connection
requests (if the phone
has such a configuration
setting).
•
(For modems) The
cellular modem is a
supported device and is
securely connected to
the X Series unit.
If all the configurations are
correct, retry the operation.