ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 44
C
HAPTER
1
M
AINTENANCE
T
ESTS
38
9650-001356-01 Rev. D
20.10 From the NIBP Test control panel, select
Open Valve to release the pressure.
20.11 From the X Series NIBP Test control
panel, use the navigation keys to select:
•
Close Valve
•
Turn Pump On
20.12 Monitor the pressure reported by the
external pressure standard.
20.13 Once the pressure reaches the target
pressure of 150 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.14 Allow the system to stabilize for at least
10 seconds.
20.15 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
o
o
20.16 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
Reading should be within
4mmHg.
o
o
20.17 From the NIBP Test control panel, select
Open Valve to release the pressure.
20.18 From the X Series NIBP Test control
panel, use the navigation keys to select:
•
Close Valve
•
Turn Pump On
20.19 Monitor the pressure reported by the
external pressure standard.
20.20 Once the pressure reaches the target
pressure of 250 +/- 5 mmHg, use the
navigation keys to select Turn Pump Off.
20.21 Allow the system to stabilize for at least
10 seconds.
20.22 Verify that the primary transducer agrees
with the reading on the NIBP simulator.
Reading should be within
3mmHg.
o
o
20.23 Verify that the safety transducer agrees
with the reading on the NIBP simulator.
Reading should be within
4mmHg.
o
o
20.24 From the NIBP Test control panel, select
Open Valve to release the pressure.
Do This...
Observe This...
Pass/Fail