beautypg.com

0 leakage current test, Leakage current test – ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 27

background image

Maintenance Tests

9650-001356-01 Rev. D

ZOLL X Series Service Manual

21

7.0 Leakage Current Test

Equipment

See the manufacturer’s instructions or supplied specifications for the leakage tester you use.

Setup

See the manufacturer’s instructions or supplied specifications for the leakage tester you use.
Repeat leakage test with accessories: MFC, external paddles, internal paddles, and anterior/
posterior paddles. Perform these tests at the line-power voltage and frequency used in your
installation.

Maximum Leakage Acceptance Limits

Patient Leakage

Mains on Applied Part

Normal

Single Fault

Single Fault

ECG

10

50

100

MFC

10

50

100

Normal

Single Fault

Earth Leakage

2500

5000