Safety precautions, Modules, Safety precautions modules – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 68
![background image](/manuals/355966/68/background.png)
C
HAPTER
3
D
ISASSEMBLY
P
ROCEDURES
62
9650-001356-01 Rev. D
•
Kapton tape
•
3M copper adhesive tape, or equivalent
•
Large diagonal cutters
Safety Precautions
WARNING!
SHOCK HAZARD!
Caution
TAKE THE NECESSARY PRECAUTIONS TO GUARD AGAINST SHOCK OR INJURY
BEFORE YOU CONDUCT DEFIBRILLATOR TESTS OR REPAIRS.
•
Only properly trained technicians should service the unit.
•
The unit can contain deadly voltages even if the unit is turned off.
•
Make sure to discharge the unit before working with it.
•
Make sure you take the necessary precautions when working with static sensitive units. For
example, you must wear a conductive wrist strap (which touches your skin) connected to a
grounding mat and to the earth ground. You must remove the wrist strap when you discharge
high voltage or when you are working on energized equipment.
Modules
The X Series unit contains the following ten (10) modules or assemblies:
•
Printer
•
Front Enclosure
•
Display
•
NIBP
•
EtCO
2
(Oridion
®
)
•
Monitor Board
•
Capacitor Bank
•
Defibrillator Board
•
SpO
2
•
CP Carrier Module
These modules can be purchased as replacement parts from the ZOLL Technical Support
Department. See Chapter 4, "Replacement Parts" for information on ordering replacement
parts.