0 treatment report test, Treatment report test – ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 35
Maintenance Tests
9650-001356-01 Rev. D
ZOLL X Series Service Manual
29
13.0 Treatment Report Test
Equipment
Fluke Impulse 7000 DP Analyzer (or equivalent)
Do this...
Observe this...
Pass/Fail
13.1 Connect the multifunction cable (MFC) to the
Analyzer, and turn on the Analyzer.
13.2 Turn on the X Series unit, and press the
following quick access keys:
a)
b)
c)
The Clear Log window
displays.
13.3 Select Yes.
The unit displays a clearing
the log message.
o
o
13.4 Press the CHARGE button, select 200 joules,
then press the CHARGE button again.
The unit starts to charge, a
charging tone sounds, and the
Shock button illuminates.
o
o
13.5 Press the Shock button.
The unit discharges the shock.
o
o
13.6 Wait 20 seconds, press the Treatment quick
access key(
), and select the first code
marker.
13.7 Wait 15 seconds and turn off the unit, then wait
2 minutes and turn on the unit.
13.8 On the X Series unit, press the following quick
access keys:
a)
b)
c)
The Treatment Summary
window displays.
o
o
13.9 Highlight the second event and press Select.
Verify that a green check mark
displays next to the file.
o
o
13.10 Select Print Treatment Summary.
Verify that the recorder prints
out the defib event and the
code marker snapshot.
o
o
13.11 Turn off the X Series unit.
LOG
R
LOG