ZOLL X Series Monitor Defibrillator Rev D User Manual
Page 57
X Series Error Messages
9650-001356-01 Rev. D
X Series Service Manual
51
CHECK COMM
CONFIGURATION
A configuration error
prevented successful
communication.
To correct the problem,
verify the following:
•
The reports
configuration contains
valid values for the
Server, Port, Customer
ID, and Password fields.
•
(Cellular modem) You
are using a valid cell
provider configuration
(including call number)
and the provider name is
specified as the USB
cell modem provider.
•
(Bluetooth DUN phone)
A valid cell provider
configuration (including
call number) is present
and the provider name is
associated with the DUN
phone pairing.
CO2 BAROMETRIC OUT
OF RANGE
Altitude is too high for
module.
Bring device into specified
altitude range. If problem
persists, replace the CO
2
module.
CO2 CAL FAILURE
Calibration failed.
Retry calibration or
continue with old
calibration. Replace the
CO
2
module.
CO2 CAL REMINDER
Calibration reminder.
Calibrate the CO
2
module
soon.
CO2 CAL REQUIRED
Calibration due.
Calibrate the CO
2
module.
CO2 CHECK FLOW
Exhaust port blocked.
Check for exhaust port
occlusion. If problem
persists, replace the CO
2
module.
CO2 COM ERROR CODE
Communication failure.
Check communication
cables. If problem persists,
replace the CO
2
module.
CO2 DEVICE FAILURE
Malfunction or device failure.
Replace the CO
2
module.
Error Message
Explanation
Technical Action