beautypg.com

ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 74

background image

C

HAPTER

3

D

ISASSEMBLY

P

ROCEDURES

68

9650-001356-01 Rev. D

6

Disconnect the battery
communication cable (yellow,
orange, brown, red)
connector from the monitor
board.

7

Disconnect the monitor/
defibrillator flex cable and
MFC cable connectors from
the monitor board.

Note: Be careful to pull the
connector straight out.
Rocking the connector or
disconnecting it at an angle
may result in bent pins.

8

Disconnect the three power
cable connectors (battery
(black), AUX (white), and
defibrillator (red)).