beautypg.com

ZOLL X Series Monitor Defibrillator Rev D User Manual

Page 87

background image

Removing the Monitor Board and SpO2 Modules

9650-001356-01 Rev. D

X Series Service Manual

81

2

Disconnect the SpO

2

ground

cable.

3

Remove 4 screws on the left
side in the illustrated areas.

Note: During reassembly,
torque screws to 6 in-lbs.

4

Turn the unit over.

5

Remove 3 screws on the left
side of the unit in the
illustrated areas.

6

Turn the unit over.